Determination of Eight Estrogens in Milk by UHPLC and the

Transcript Of Determination of Eight Estrogens in Milk by UHPLC and the
Determination of Eight Estrogens in Milk by UHPLC and the Agilent 6495 Triple Quadrupole Mass Spectrometer
Authors
Dan-Hui Dorothy Yang, Agilent Technologies Inc., Santa Clara, California, USA
Jian-Zhong Li, Agilent Technologies, Inc., Beijing, P.R. China
Bernhard Wuest, Agilent Technologies R&D and Marketing GmbH & Co. KG, Waldbronn, Germany
Application Note
Abstract
This Application Note demonstrates a complete method to rapidly and precisely determine estrogens in milk with the Agilent 6495 Triple Quadrupole Mass Spectrometer coupled with the Agilent 1290 Infinity UHPLC. Using dynamic multiple reaction monitoring (DMRM) in negative ion mode, quantitation of estrogens was achieved with a simple QuEChERS procedure without any derivatization. The lower limits of quantitation (LLOQ) for all estrogens in milk were less than 0.1 μg/kg, far below the regulatory requirement, which is set at 1 μg/kg4.
Introduction
Human exposure to estrogens through the consumption of cow milk and other dairy products is an area of increasing concern due to the considerable amounts of female sex hormones they contain. On modern dairy farms, cows are milked approximately 300 days a year, and are pregnant for the majority of that time. The milk produced by a pregnant cow is higher in estrogens. This leads to an increased exposure of these hormones when consumed by the public.
The presence of estrogens in milk impacts the development of infants and children, causing early maturation in girls and delayed maturation in boys1. Regular exposure to high levels of estrogens has been linked to hormone-sensitive cancers, such as breast cancer and prostate cancer2,3. Routine determination of estrogens in milk or milk powder is required by many governmental agencies4,5.
Generally accepted methods for analyzing estrogen in milk employ a traditional solvent extraction followed by solid phase extraction (SPE)1. Although they have been widely accepted, these traditional methods are labor-intensive, use a large amount of solvent, and generate a high volume of waste. The use of highlysensitive instrumentation avoids the need of enrichment of a large sample size or analyte derivatization. We have developed a method using an Agilent QuEChERS kit6 in combination with the highly sensitive Agilent 6495 Triple Quadrupole Mass Spectrometer to quantify estrogens in
OH
OH
H
HH HO
17-a-Estradiol
H
HH
HO
HO
17-b-Estradiol
HO Dienestrol
OH HO
O
H HH
HO Estrone
OH
H
OH
HH
Estriol OH
H HH
Ethynyl estradiol
HO
H3C
OH
OH
HO
Diethylstilbestrol
CH3 Hexestrol
Figure 1. Structures of estrogens detected with the LC/MS/MS method.
milk. Several modifications to the triple quadrupole mass spectrometer have resulted in better analytical performance. These modifications include a new mass filter one optics for increased precursor ion transmission, an improved curved and tapered collision cell for enhanced MS/MS spectral fidelity, a new ion detector operating at dynode accelerating voltages of up to 20 kV, and an optimized autotune for speed and sensitivity. This method provides significant advantages
over the traditional methods, including minimal labor, high recovery, high precision, and limited solvent usage. We have achieved LLOQs far below regulatory requirements for the following eight estrogens: estriol, estradiol (two isomers: 17-a-estradiol and 17-b-estradiol), estrone, dienestrol, ethynyl estradiol, diethylstilbestrol, and hexestrol. Figure 1 shows the structure of eight estrogens detected with the method.
2
Experimental
Reagents and chemicals
All reagents and solvents were HPLC or analytical grade. Methanol and acetonitrile were purchased from Honeywell (Catalog number 230-4 and 015-4 respectively). Ultrapure water was obtained from a Milli-Q Integral system equipped with LC-Pak Polisher and a 0.22-μm membrane point-of-use cartridge (Millipak). Ammonium fluoride was purchased from Fluka (338869-25 g). A 5 M stock solution was made by dissolving ammonium fluoride in Milli-Q water. Estrogen standards were obtained from the Chinese Academy of Inspection and Quarantine at concentrations of 10 ppm, 1 ppm, and 100 ppb in acetonitrile.
Instrumentation and conditions
• Agilent 1290 Infinity Binary Pump (G4220A)
• Agilent 1290 Infinity Standard Autosampler (G4226A) and sample cooler (G1330B)
• Agilent 1290 Infinity Thermostatted Column Compartment (G1316C)
UHPLC conditions are listed in Table 1.
MS detection
• Agilent 6495 Triple Quadrupole Mass Spectrometer with Agilent JetStream electrospray ionization source
Software
• Agilent MassHunter data acquisition for triple quadruple mass spectrometer, Version B.07.00
• Agilent MassHunter Qualitative Software, Version B.06.0.633.10 SP1
• Agilent MassHunter Quantitative Software, Version B.06.00.388.00
MS conditions are listed in Table 1. Table 2 lists the precursor ions/product ions and their corresponding optimized collision energy values for eight estrogens.
Caliper LifeSciences Turbo Vap LV Eppendorf Centrifuge 5451 R Heraeus Labofuge 400 R
Table 1. Chromatographic conditions.
UHPLC conditions
Column
Agilent Poroshell 120 Phenyl-Hexyl 2.1 × 100 mm, 2.7 μm (p/n 695775-912)
Column temperature
35 °C
Injection volume
5 μL
Speed
Draw 100 μL/min; Eject 200 μL/min
Autosampler temperature 6 °C
Needle wash
20 seconds (80 % MEOH/20 % water)
Mobile phase
A) Water with 0.4 mM ammonium fluoride
B) Methanol:acetonitrile (1:1 v/v)
Flow rate
0.4 mL/min
Gradient program
Time
B %
0
15
0.5
15
2.5
40
6.0
65
7.0
95
9.0
95
9.1
15
MS conditions
Stop time Post time
10 minutes 2 minutes
Ion mode Drying gas temperature Drying gas flow Sheath gas temperature Sheath gas flow Nebulizer pressure Capillary voltage Nozzle voltage Delta EMV LPF RF HPF RF MS1 and MS2 resolution
Negative 200 16 390 12 35 3,000 (neg) 300 (neg) 400 (neg) 60 (neg) 100 (neg) Unit
3
Sample preparation
The sample preparation procedure includes extraction/partitioning and dispersive SPE cleanup. Two percent organic milk was purchased from a local grocery store. Ten-gram milk samples were fortified with the appropriate amount of estrogens. After a brief vortex, 10 mL of acetonitrile was added and the tube was vortexed for 30 seconds. One pouch of Agilent Bond Elute extraction/ partitioning reagent (p/n 5982-5755) was then added to each tube, followed by vigorous shaking for 1 minute. The tubes were then centrifuged for 5 minutes at 3,500 rpm.
A 6 mL amount of supernatant acetonitrile was transferred to an Agilent Bond Elut QuEChERS dispersive SPE 15-mL tube for animal tissue (p/n 6982-4956). The tubes were vortexed for 1 minute and centrifuged at 3,500 rpm for 5 minutes. Then, 4 mL of liquid layer was transferred to a glass test tube and dried at 40 °C under constant nitrogen flow. The samples were reconstituted with 2 mL of 50 % methanol water. After reconstitution, the samples were transferred to a 2-mL Eppendorf tube and centrifuged at 14,000 rpm for 7 minutes at 4 °C. The supernatant clear solution was then transferred to an HPLC vial for injection. Figure 2 shows the flow chart of the QuEChERS procedure for the determination of estrogens in milk.
Table 2. Optimum precursor/product ions and their corresponding collision energy (CE) values. Dwell time depends on the number of MRMs in a dynamic MRM method. Min/max dwell time of 31.14 ms and 224.11 ms were defined with cycle time 450 ms.
Compound name
17-a-estradiol 17-a-estradiol 17-b-estradiol 17-b-estradiol Dienestrol Dienestrol Diethylstilbestrol Diethylstibestrol Estriol Estriol Estrone Estrone Ethynyl estradiol Ethynyl estradiol Hexestrol Hexestrol
Precursor ion 271.1 271.1 271.15 271.15 265.1 265.1 267.1 267.1 287.1 287.1 269.1 269.1 295.2 295.2 269.1 269.1
Product ion 145 183 183.1 145.1 249 93 251 237 171 145 159 145 269 145 133 119
RT (min) 5.2 5.2 5.4 5.4 6.17 6.17 5.94 5.94 3.53 3.53 5.81 5.81 5.55 5.55 6.12 6.12
Delta RT 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1
Fragmentor 380 380 380 380 380 380 380 380 380 380 380 380 380 380 380 380
CE CAV Polarity
45 3
Negative
45 3
Negative
45 3
Negative
45 3
Negative
25 3
Negative
30 3
Negative
27 3
Negative
30 3
Negative
43 3
Negative
50 3
Negative
42 3
Negative
45 3
Negative
35 3
Negative
45 3
Negative
15 3
Negative
43 3
Negative
Weigh 10 g of milk sample (± 0.05 g) into a 50-mL centrifuge tube
Spike estrogens with different working stock solutions 5–20 µL, vortex 30 seconds
Add 10 mL acetonitrile, vortex 30 seconds
Add one pouch of Agilent Bond Elut QuEChERS (5982-5755) and shake vigorously for 1 minute
Centrifuge at 3,500 rpm for 5 minutes
Transfer 6 mL of the ACN layer to a an Agilent Bond Elut QuEChERS dispersive SPE 15-mL tube (6982-4956 for animal tissue)
Vortex for 1 minute and centrifuge at 3,500 rpm for 5 minutes
Transfer 4 mL of the ACN layer to a glass test tube and dry at 40 °C with N2
Reconstitute with 2 mL of 50 % MeOH in H2O and centrifuge at 14,000 rpm at 4 °C for 7 minutes
Transfer clear solution into an HPLC vial for injection Figure 2. Flow chart of QuEChERS procedure for the determination of estrogens in milk.
4
Results and Discussion
Instrument detection limit (IDL) of estrogens in solvent
To test the sensitivity of the instrument, and the feasibility of estrogen detection, eight estrogens were diluted in 50 % methanol water from 0.2 ppt to 100 ppb from a working solution of 1 ppm at 12 levels. IDL refers to the minimum amount of analyte required to produce a signal that is statistically distinguishable from background noise with a given confidence level. This approach helps avoid ambiguity related to the variation in the chemical noise and the different ways in which signal-to-noise ratios (S/N) are determined7. The LLOQs for these compounds range from 5 to 10 ppt, and the IDL ranges roughly from 1.2 to 3.7 ppt as listed in Table 3. The linear dynamic range covers 10 ppt to 100 ppb with a correlation coefficient of R2 > 0.994 for all analytes. These results demonstrate the high sensitivity of the 6495 Triple Quadrupole Mass Spectrometer for estrogen detection.
Linearity and limit of quantitation (LOQ) of estrogens in milk
The linear calibration range for all estrogens in milk was from 0.1 µg/kg to 10 µg/kg. Appropriate amounts of estrogens were spiked into 10 g of milk. The spike levels were 0.1, 0.2, 0.5, 1, 2, 5, and 10 µg/kg from working stock solutions of 100 ppb to 10 ppm (5–20 µL). The sample preparation procedure is outlined in Figure 2. The calibration curves were generated by plotting the relative response (peak area) versus the concentration of estrogens. The LOQ should be less than the lowest spike concentration tested at 0.1 µg/kg, which is far below the regulatory requirement of 1 µg/kg. Figure 3 shows the normalized chromatogram of 0.1 µg/kg spiked estrogens in milk and their associated retention time. Figure 4 shows the overlay of a blank milk sample and the 0.1 µg/kg spiked milk sample.
Table 3. IDL and LLOQ for eight estrogens in solvent, demonstrating highly sensitive detection in negative ion mode.
Analytes Estriol 17-a-Estradiol 17-b-Estradiol Ethynyl estradiol Estrone Diethylstilbestrol Hexestrol Dienestrol
IDL (ppt) 1.78 2.65 3.60 3.69 1.50 2.02 1.74 1.20
LLOQ (ppt) 5.0 10.0 10.0 10.0 5.0 10.0 5.0 5.0
Counts (%)
×102 1.1
Estriol 1.0 0.9
Ethynyl estradiol Diethylstilbestrol
17-b-Estradiol
Hexestrol
17-a-Estradiol
Estrone
Dienestrol
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
3.2 3.4 3.6 3.8 4 4.2 4.4 4.6 4.8 5 5.2 5.4 5.6 5.8 6 6.2 6.4 6.6 Acquisition time (min)
Figure 3. Normalized chromatogram of the 0.1 µg/kg spiked estrogens in milk.
Counts (%)
×103 3.2
3 2.8 2.6 2.4 2.2
2 1.8 1.6 1.4 1.2
1 0.8 0.6 0.4 0.2
0
3.2 3.4 3.6 3.8 4 4.2 4.4 4.6 4.8 5 5.2 5.4 5.6 5.8 6 Acquisition time (min)
Figure 4. Overlay of 0.1 µg/kg estrogens spiked in milk (red) versus milk blank (black).
6.2 6.4 6.6
5
The milk used for the blank and spiked injections may contain a very small amount of estrone, but its quantity is far less than the 0.1 µg/kg spiked concentration. Thus, this should not be a concern. Table 4 shows the regression equation and correlation coefficient (R2) for eight estrogens from 0.1 µg/kg to 10 µg/kg at seven calibration levels. Linear regression fit of 1/x was applied. Results indicated excellent linearity for all analytes.
Recovery and reproducibility
The recovery and reproducibility were evaluated by fortifying estrogen standards in milk at levels of 0.2 µg/kg, 1 µg/kg, and 5 µg/kg. The analysis included six spiked replicates at each concentration. The QuEChERS procedure recovery was calculated as the ratio of the area for the pre-extraction spike (average of six replicates) and post-extraction spike at the same final concentration. Table 5 shows the recovery and reproducibility (listed as RSD % for six replicates). Results indicate that all eight estrogens gave good recoveries, especially for more polar compounds. Precision values were excellent for all analytes at three levels, primarily < 10 %.
Table 4. Linearity of estrogens in milk.
Analytes Estriol 17-a-Estradiol 17-b-Estradiol Ethynyl estradiol Estrone Diethylstilbestrol Hexestrol Dienestrol
Regression equation Y = 23308.11X - 235.73 Y = 11519.06X - 244.56 Y = 16435.07X - 324.27 Y = 13872.40X - 281.85 Y = 113277.15X - 1685.45 Y = 212726.81X + 572.60 Y = 223738X - 8467.75 Y = 169692X - 1949.39
R2 0.9988 0.9973 0.9985 0.9984 0.9987 0.9985 0.9981 0.9995
Table 5. Recovery and reproducibility of estrogens spiked in milk.
Analytes Estriol 17-a-Estradiol 17-b-Estradiol Ethynyl estradiol Estrone Diethylstilbestrol Hexestrol Dienestrol
0.2 µg/kg (n = 6)
Recovery RSD
92.9
5.2
92.2
7.8
92.9
3.3
90.2
4.8
94.0
2.9
83.5
3.7
86.1
7.0
82.9
3.6
1 µg/kg (n = 6)
Recovery RSD
102.8
4.9
99.4
8.5
106.0
5.8
104.9
4.7
101.7
5.6
98.6
3.7
98.4
3.1
92.8
2.0
5 µg/kg (n = 6)
Recovery RSD
85.8
4.8
93.4
8.5
96.9
8.8
94.0
8.4
96.3
6.8
91.0
14.6
91.8
13.0
85.3
9.5
6
Conclusion
The highly sensitive Agilent 6495 Triple Quadrupole Mass Spectrometer is a potential tool for estrogen measurement in food matrixes where, by regulation, MRLs are set low. The advanced instrumentation also allows the use of simpler sample preparation procedures, as significant enrichment or derivatization is unnecessary. The Agilent Bond Elut Buffered QuEChERS method is easier to implement and saves solvent when compared to other methods such as typical LLE and SPE. The recovery and precision of the method, based on the matrix-spiked standards, were excellent for multi-residue estrogen determination in milk. The impurities and observed matrix effect were minimal for the milk blank, which did not interfere with the quantitation of the analytes at 10x less than the MRL. This method can be applied to the measurement of estrogens in other similar matrixes, such as milk powder.
References
1. Maruyama, K.; Oshima, T.; Ohyama, K. Exposure to exogenous estrogen through intake of commercial milk produced from pregnant cows. Pediatrics International 2010, 52(1) pp 33-38.
2. Candyce, H., et al. High- and LowFat Dairy Intake, Recurrence, and Mortality After Breast Cancer Diagnosis. J. of the Nat. Cancer Inst. 2013, 105(22), pp 1759-60.
3. Qin, L.Q., et al. Estrogen: one of the risk factors in milk for prostate cancer. Med. Hypotheses 2004, 62(1), pp 133-42.
4. GB/T 22992-2008, P.R. China.
5. 96/23/EC, European Union, Council Directive.
6. Stubbing, G. and Bigwood, T. The development and validation of a multiclass liquid chromatography tandem mass spectrometry (LC/MS/ MS) procedure for the determination of veterinary drug residue in animal tissue using QuEChERS (Quick, Easy, Cheap, Effevtive, Rugged and Safe) approach. Analytica Chimica Acta 2009, 637, pp 68-78.
7. Parra, N.P.; Taylor, L. Why Instrument Detection Limit (IDL) Is a Better Metric for Determining the Sensitivity of Triple Quadrupole LC/MS Systems, Agilent Technologies Technical Overview, publication number 5991-4089EN, 2014.
7
www.agilent.com/chem
This information is subject to change without notice.
© Agilent Technologies, Inc., 2014 Published in the USA, June 24, 2014 5991-4686EN
Authors
Dan-Hui Dorothy Yang, Agilent Technologies Inc., Santa Clara, California, USA
Jian-Zhong Li, Agilent Technologies, Inc., Beijing, P.R. China
Bernhard Wuest, Agilent Technologies R&D and Marketing GmbH & Co. KG, Waldbronn, Germany
Application Note
Abstract
This Application Note demonstrates a complete method to rapidly and precisely determine estrogens in milk with the Agilent 6495 Triple Quadrupole Mass Spectrometer coupled with the Agilent 1290 Infinity UHPLC. Using dynamic multiple reaction monitoring (DMRM) in negative ion mode, quantitation of estrogens was achieved with a simple QuEChERS procedure without any derivatization. The lower limits of quantitation (LLOQ) for all estrogens in milk were less than 0.1 μg/kg, far below the regulatory requirement, which is set at 1 μg/kg4.
Introduction
Human exposure to estrogens through the consumption of cow milk and other dairy products is an area of increasing concern due to the considerable amounts of female sex hormones they contain. On modern dairy farms, cows are milked approximately 300 days a year, and are pregnant for the majority of that time. The milk produced by a pregnant cow is higher in estrogens. This leads to an increased exposure of these hormones when consumed by the public.
The presence of estrogens in milk impacts the development of infants and children, causing early maturation in girls and delayed maturation in boys1. Regular exposure to high levels of estrogens has been linked to hormone-sensitive cancers, such as breast cancer and prostate cancer2,3. Routine determination of estrogens in milk or milk powder is required by many governmental agencies4,5.
Generally accepted methods for analyzing estrogen in milk employ a traditional solvent extraction followed by solid phase extraction (SPE)1. Although they have been widely accepted, these traditional methods are labor-intensive, use a large amount of solvent, and generate a high volume of waste. The use of highlysensitive instrumentation avoids the need of enrichment of a large sample size or analyte derivatization. We have developed a method using an Agilent QuEChERS kit6 in combination with the highly sensitive Agilent 6495 Triple Quadrupole Mass Spectrometer to quantify estrogens in
OH
OH
H
HH HO
17-a-Estradiol
H
HH
HO
HO
17-b-Estradiol
HO Dienestrol
OH HO
O
H HH
HO Estrone
OH
H
OH
HH
Estriol OH
H HH
Ethynyl estradiol
HO
H3C
OH
OH
HO
Diethylstilbestrol
CH3 Hexestrol
Figure 1. Structures of estrogens detected with the LC/MS/MS method.
milk. Several modifications to the triple quadrupole mass spectrometer have resulted in better analytical performance. These modifications include a new mass filter one optics for increased precursor ion transmission, an improved curved and tapered collision cell for enhanced MS/MS spectral fidelity, a new ion detector operating at dynode accelerating voltages of up to 20 kV, and an optimized autotune for speed and sensitivity. This method provides significant advantages
over the traditional methods, including minimal labor, high recovery, high precision, and limited solvent usage. We have achieved LLOQs far below regulatory requirements for the following eight estrogens: estriol, estradiol (two isomers: 17-a-estradiol and 17-b-estradiol), estrone, dienestrol, ethynyl estradiol, diethylstilbestrol, and hexestrol. Figure 1 shows the structure of eight estrogens detected with the method.
2
Experimental
Reagents and chemicals
All reagents and solvents were HPLC or analytical grade. Methanol and acetonitrile were purchased from Honeywell (Catalog number 230-4 and 015-4 respectively). Ultrapure water was obtained from a Milli-Q Integral system equipped with LC-Pak Polisher and a 0.22-μm membrane point-of-use cartridge (Millipak). Ammonium fluoride was purchased from Fluka (338869-25 g). A 5 M stock solution was made by dissolving ammonium fluoride in Milli-Q water. Estrogen standards were obtained from the Chinese Academy of Inspection and Quarantine at concentrations of 10 ppm, 1 ppm, and 100 ppb in acetonitrile.
Instrumentation and conditions
• Agilent 1290 Infinity Binary Pump (G4220A)
• Agilent 1290 Infinity Standard Autosampler (G4226A) and sample cooler (G1330B)
• Agilent 1290 Infinity Thermostatted Column Compartment (G1316C)
UHPLC conditions are listed in Table 1.
MS detection
• Agilent 6495 Triple Quadrupole Mass Spectrometer with Agilent JetStream electrospray ionization source
Software
• Agilent MassHunter data acquisition for triple quadruple mass spectrometer, Version B.07.00
• Agilent MassHunter Qualitative Software, Version B.06.0.633.10 SP1
• Agilent MassHunter Quantitative Software, Version B.06.00.388.00
MS conditions are listed in Table 1. Table 2 lists the precursor ions/product ions and their corresponding optimized collision energy values for eight estrogens.
Caliper LifeSciences Turbo Vap LV Eppendorf Centrifuge 5451 R Heraeus Labofuge 400 R
Table 1. Chromatographic conditions.
UHPLC conditions
Column
Agilent Poroshell 120 Phenyl-Hexyl 2.1 × 100 mm, 2.7 μm (p/n 695775-912)
Column temperature
35 °C
Injection volume
5 μL
Speed
Draw 100 μL/min; Eject 200 μL/min
Autosampler temperature 6 °C
Needle wash
20 seconds (80 % MEOH/20 % water)
Mobile phase
A) Water with 0.4 mM ammonium fluoride
B) Methanol:acetonitrile (1:1 v/v)
Flow rate
0.4 mL/min
Gradient program
Time
B %
0
15
0.5
15
2.5
40
6.0
65
7.0
95
9.0
95
9.1
15
MS conditions
Stop time Post time
10 minutes 2 minutes
Ion mode Drying gas temperature Drying gas flow Sheath gas temperature Sheath gas flow Nebulizer pressure Capillary voltage Nozzle voltage Delta EMV LPF RF HPF RF MS1 and MS2 resolution
Negative 200 16 390 12 35 3,000 (neg) 300 (neg) 400 (neg) 60 (neg) 100 (neg) Unit
3
Sample preparation
The sample preparation procedure includes extraction/partitioning and dispersive SPE cleanup. Two percent organic milk was purchased from a local grocery store. Ten-gram milk samples were fortified with the appropriate amount of estrogens. After a brief vortex, 10 mL of acetonitrile was added and the tube was vortexed for 30 seconds. One pouch of Agilent Bond Elute extraction/ partitioning reagent (p/n 5982-5755) was then added to each tube, followed by vigorous shaking for 1 minute. The tubes were then centrifuged for 5 minutes at 3,500 rpm.
A 6 mL amount of supernatant acetonitrile was transferred to an Agilent Bond Elut QuEChERS dispersive SPE 15-mL tube for animal tissue (p/n 6982-4956). The tubes were vortexed for 1 minute and centrifuged at 3,500 rpm for 5 minutes. Then, 4 mL of liquid layer was transferred to a glass test tube and dried at 40 °C under constant nitrogen flow. The samples were reconstituted with 2 mL of 50 % methanol water. After reconstitution, the samples were transferred to a 2-mL Eppendorf tube and centrifuged at 14,000 rpm for 7 minutes at 4 °C. The supernatant clear solution was then transferred to an HPLC vial for injection. Figure 2 shows the flow chart of the QuEChERS procedure for the determination of estrogens in milk.
Table 2. Optimum precursor/product ions and their corresponding collision energy (CE) values. Dwell time depends on the number of MRMs in a dynamic MRM method. Min/max dwell time of 31.14 ms and 224.11 ms were defined with cycle time 450 ms.
Compound name
17-a-estradiol 17-a-estradiol 17-b-estradiol 17-b-estradiol Dienestrol Dienestrol Diethylstilbestrol Diethylstibestrol Estriol Estriol Estrone Estrone Ethynyl estradiol Ethynyl estradiol Hexestrol Hexestrol
Precursor ion 271.1 271.1 271.15 271.15 265.1 265.1 267.1 267.1 287.1 287.1 269.1 269.1 295.2 295.2 269.1 269.1
Product ion 145 183 183.1 145.1 249 93 251 237 171 145 159 145 269 145 133 119
RT (min) 5.2 5.2 5.4 5.4 6.17 6.17 5.94 5.94 3.53 3.53 5.81 5.81 5.55 5.55 6.12 6.12
Delta RT 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1
Fragmentor 380 380 380 380 380 380 380 380 380 380 380 380 380 380 380 380
CE CAV Polarity
45 3
Negative
45 3
Negative
45 3
Negative
45 3
Negative
25 3
Negative
30 3
Negative
27 3
Negative
30 3
Negative
43 3
Negative
50 3
Negative
42 3
Negative
45 3
Negative
35 3
Negative
45 3
Negative
15 3
Negative
43 3
Negative
Weigh 10 g of milk sample (± 0.05 g) into a 50-mL centrifuge tube
Spike estrogens with different working stock solutions 5–20 µL, vortex 30 seconds
Add 10 mL acetonitrile, vortex 30 seconds
Add one pouch of Agilent Bond Elut QuEChERS (5982-5755) and shake vigorously for 1 minute
Centrifuge at 3,500 rpm for 5 minutes
Transfer 6 mL of the ACN layer to a an Agilent Bond Elut QuEChERS dispersive SPE 15-mL tube (6982-4956 for animal tissue)
Vortex for 1 minute and centrifuge at 3,500 rpm for 5 minutes
Transfer 4 mL of the ACN layer to a glass test tube and dry at 40 °C with N2
Reconstitute with 2 mL of 50 % MeOH in H2O and centrifuge at 14,000 rpm at 4 °C for 7 minutes
Transfer clear solution into an HPLC vial for injection Figure 2. Flow chart of QuEChERS procedure for the determination of estrogens in milk.
4
Results and Discussion
Instrument detection limit (IDL) of estrogens in solvent
To test the sensitivity of the instrument, and the feasibility of estrogen detection, eight estrogens were diluted in 50 % methanol water from 0.2 ppt to 100 ppb from a working solution of 1 ppm at 12 levels. IDL refers to the minimum amount of analyte required to produce a signal that is statistically distinguishable from background noise with a given confidence level. This approach helps avoid ambiguity related to the variation in the chemical noise and the different ways in which signal-to-noise ratios (S/N) are determined7. The LLOQs for these compounds range from 5 to 10 ppt, and the IDL ranges roughly from 1.2 to 3.7 ppt as listed in Table 3. The linear dynamic range covers 10 ppt to 100 ppb with a correlation coefficient of R2 > 0.994 for all analytes. These results demonstrate the high sensitivity of the 6495 Triple Quadrupole Mass Spectrometer for estrogen detection.
Linearity and limit of quantitation (LOQ) of estrogens in milk
The linear calibration range for all estrogens in milk was from 0.1 µg/kg to 10 µg/kg. Appropriate amounts of estrogens were spiked into 10 g of milk. The spike levels were 0.1, 0.2, 0.5, 1, 2, 5, and 10 µg/kg from working stock solutions of 100 ppb to 10 ppm (5–20 µL). The sample preparation procedure is outlined in Figure 2. The calibration curves were generated by plotting the relative response (peak area) versus the concentration of estrogens. The LOQ should be less than the lowest spike concentration tested at 0.1 µg/kg, which is far below the regulatory requirement of 1 µg/kg. Figure 3 shows the normalized chromatogram of 0.1 µg/kg spiked estrogens in milk and their associated retention time. Figure 4 shows the overlay of a blank milk sample and the 0.1 µg/kg spiked milk sample.
Table 3. IDL and LLOQ for eight estrogens in solvent, demonstrating highly sensitive detection in negative ion mode.
Analytes Estriol 17-a-Estradiol 17-b-Estradiol Ethynyl estradiol Estrone Diethylstilbestrol Hexestrol Dienestrol
IDL (ppt) 1.78 2.65 3.60 3.69 1.50 2.02 1.74 1.20
LLOQ (ppt) 5.0 10.0 10.0 10.0 5.0 10.0 5.0 5.0
Counts (%)
×102 1.1
Estriol 1.0 0.9
Ethynyl estradiol Diethylstilbestrol
17-b-Estradiol
Hexestrol
17-a-Estradiol
Estrone
Dienestrol
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
3.2 3.4 3.6 3.8 4 4.2 4.4 4.6 4.8 5 5.2 5.4 5.6 5.8 6 6.2 6.4 6.6 Acquisition time (min)
Figure 3. Normalized chromatogram of the 0.1 µg/kg spiked estrogens in milk.
Counts (%)
×103 3.2
3 2.8 2.6 2.4 2.2
2 1.8 1.6 1.4 1.2
1 0.8 0.6 0.4 0.2
0
3.2 3.4 3.6 3.8 4 4.2 4.4 4.6 4.8 5 5.2 5.4 5.6 5.8 6 Acquisition time (min)
Figure 4. Overlay of 0.1 µg/kg estrogens spiked in milk (red) versus milk blank (black).
6.2 6.4 6.6
5
The milk used for the blank and spiked injections may contain a very small amount of estrone, but its quantity is far less than the 0.1 µg/kg spiked concentration. Thus, this should not be a concern. Table 4 shows the regression equation and correlation coefficient (R2) for eight estrogens from 0.1 µg/kg to 10 µg/kg at seven calibration levels. Linear regression fit of 1/x was applied. Results indicated excellent linearity for all analytes.
Recovery and reproducibility
The recovery and reproducibility were evaluated by fortifying estrogen standards in milk at levels of 0.2 µg/kg, 1 µg/kg, and 5 µg/kg. The analysis included six spiked replicates at each concentration. The QuEChERS procedure recovery was calculated as the ratio of the area for the pre-extraction spike (average of six replicates) and post-extraction spike at the same final concentration. Table 5 shows the recovery and reproducibility (listed as RSD % for six replicates). Results indicate that all eight estrogens gave good recoveries, especially for more polar compounds. Precision values were excellent for all analytes at three levels, primarily < 10 %.
Table 4. Linearity of estrogens in milk.
Analytes Estriol 17-a-Estradiol 17-b-Estradiol Ethynyl estradiol Estrone Diethylstilbestrol Hexestrol Dienestrol
Regression equation Y = 23308.11X - 235.73 Y = 11519.06X - 244.56 Y = 16435.07X - 324.27 Y = 13872.40X - 281.85 Y = 113277.15X - 1685.45 Y = 212726.81X + 572.60 Y = 223738X - 8467.75 Y = 169692X - 1949.39
R2 0.9988 0.9973 0.9985 0.9984 0.9987 0.9985 0.9981 0.9995
Table 5. Recovery and reproducibility of estrogens spiked in milk.
Analytes Estriol 17-a-Estradiol 17-b-Estradiol Ethynyl estradiol Estrone Diethylstilbestrol Hexestrol Dienestrol
0.2 µg/kg (n = 6)
Recovery RSD
92.9
5.2
92.2
7.8
92.9
3.3
90.2
4.8
94.0
2.9
83.5
3.7
86.1
7.0
82.9
3.6
1 µg/kg (n = 6)
Recovery RSD
102.8
4.9
99.4
8.5
106.0
5.8
104.9
4.7
101.7
5.6
98.6
3.7
98.4
3.1
92.8
2.0
5 µg/kg (n = 6)
Recovery RSD
85.8
4.8
93.4
8.5
96.9
8.8
94.0
8.4
96.3
6.8
91.0
14.6
91.8
13.0
85.3
9.5
6
Conclusion
The highly sensitive Agilent 6495 Triple Quadrupole Mass Spectrometer is a potential tool for estrogen measurement in food matrixes where, by regulation, MRLs are set low. The advanced instrumentation also allows the use of simpler sample preparation procedures, as significant enrichment or derivatization is unnecessary. The Agilent Bond Elut Buffered QuEChERS method is easier to implement and saves solvent when compared to other methods such as typical LLE and SPE. The recovery and precision of the method, based on the matrix-spiked standards, were excellent for multi-residue estrogen determination in milk. The impurities and observed matrix effect were minimal for the milk blank, which did not interfere with the quantitation of the analytes at 10x less than the MRL. This method can be applied to the measurement of estrogens in other similar matrixes, such as milk powder.
References
1. Maruyama, K.; Oshima, T.; Ohyama, K. Exposure to exogenous estrogen through intake of commercial milk produced from pregnant cows. Pediatrics International 2010, 52(1) pp 33-38.
2. Candyce, H., et al. High- and LowFat Dairy Intake, Recurrence, and Mortality After Breast Cancer Diagnosis. J. of the Nat. Cancer Inst. 2013, 105(22), pp 1759-60.
3. Qin, L.Q., et al. Estrogen: one of the risk factors in milk for prostate cancer. Med. Hypotheses 2004, 62(1), pp 133-42.
4. GB/T 22992-2008, P.R. China.
5. 96/23/EC, European Union, Council Directive.
6. Stubbing, G. and Bigwood, T. The development and validation of a multiclass liquid chromatography tandem mass spectrometry (LC/MS/ MS) procedure for the determination of veterinary drug residue in animal tissue using QuEChERS (Quick, Easy, Cheap, Effevtive, Rugged and Safe) approach. Analytica Chimica Acta 2009, 637, pp 68-78.
7. Parra, N.P.; Taylor, L. Why Instrument Detection Limit (IDL) Is a Better Metric for Determining the Sensitivity of Triple Quadrupole LC/MS Systems, Agilent Technologies Technical Overview, publication number 5991-4089EN, 2014.
7
www.agilent.com/chem